Chlorine Analyzer, Free/Total

CL17


Chlorine Analyzer, Free/Total



Details

Industry standard, CL17 Free Chlorine Analyzer. Included with the CL17 analyzer is a 1 month supply of DPD indicator and Free Chlorine indicator and buffer solutions, an installation kit, sample conditioning kit, maintenance kit, and a user manual. *This product has not been evaluated to test for chlorine and chloramines in medical applications in the United States.

  • Accurate, Reliable Results
  • Simple, Predictable Maintenance
  • Factory Calibrated
  • EPA Compliant
  • Pump Control

Specifications

Accuracy:± 5 % or ±0.03 mg/L (ppm) as CL2, whichever is greater
Air Purge:0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting
Alarm:Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Certifications:Europe, CE Approved with:
EN 61326-1
CISPR 11
EN 61010-1
IEC 60529

North America:
UL 61010A-1
CAN/CSA C22.2 No. 1010.1-92
Communication:4-20mA Output
Communication Capabilities:4-20mA Output
Cycle Time:2.5 minutes
Dimensions Metric (H x W X D):454 mm x 314 mm x 179 mm
Display:LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line
Drain:1/2-inch ID flexible hose, hose barb
Enclosure Construction:ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Inlet:1/4-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure:1 to 5 psig (0.07 to 0.34 bar), 1.5 psi is optimum
Interferences:Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr⁶⁺ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃.
Light Source:Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
Lower Limit of Detection (LOD):0.03 mg/L (ppm)
Mounting:Wall mount
Operating Humidity:Up to 90% at 40 °C (104 °F) maximum
Operating Temperature Range:5 to 40 °C (41 to 104 °F)
Power Requirements (Amps):2.5 A
Power Requirements (Hz):50/60 Hz
Power Requirements (Voltage):100 - 115/230 V AC
Pressure Range:Inlet Pressure to Sample Conditioning: 1.5 to 75 psi (0.1 bar to 5.2 bar)
Range:0 to 5 mg/L free or total residual chlorine
Repeatability:0.01 ppm or 3.0 % , whichever is greater
Sample Flow Rate:200 to 500 mL per minute minimum required
Sample Temperature:5 to 40 °C (41 to 104 °F)
Weight:24.91 lbs. (11.3 kg)