Turbidimeter (Low Range)

1720E


Turbidimeter (Low Range)



Details

    Applications: Drinking Water, Industrial Water.

    Meets and exceeds USEPA Method 180.1 for measuring turbidity for drinking water compliance. Plug and play connection. Sensitive to fine changes in low level turbidity with a built-in bubble removal system. Sensor only. Two year warranty.

    • Accuracy
    • Simplicity
    • Data Collection and Display
    • USEPA Reporting
    • Experience

    Specifications

    Accuracy:±2% of reading or ±0.02 NTU (whichever is greater) from 0 to 40 NTU
    Accuracy 2:±5% of reading from 40 to 100 NTU (Defined according to ISO 15839.)
    Alarm:When connected to sc controller: three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated 5A resistive load at 230 Vac
    Certifications:Listed by ETL to UL 61010A-1: Certified by ETL to CSA C22.2 No. 1010.1: CE certified by Hach Company to EN 61010-1
    Communication:When connected to sc controller: MODBUS RS232/RS485, PROFIBUS, HART
    Compliance:Standard Methods 2130B, USEPA 180.1, Hach Method 8195
    Dimensions Metric (H x W X D):254 mm x 305 mm x 406 mm
    Mounting:Turbidimeter body and head assembly: wall and floor stand; sc controller: wall, pole, panel, and floor stand
    Operating Humidity:5 to 95 % non-condensing
    Operating Temperature Range:0 to 40 °C (Double Sensor System)
    0 to 50 °C (Single Sensor System)
    Output:Up to five selectable for 0-20 mA or 4-20 mA; output span programmable over any portion of the 0-100 NTU range; built into the sc Controller
    Power Requirements (Hz):50 - 60 Hz
    Power Requirements (Voltage):None
    Range:0.001 to 100 Nephelometric Turbidity Units (NTU)
    Repeatability:Better than ± 1.0 % of reading or ±0.002 NTU, whichever is greater (Defined according to ISO 15839.)
    Resolution:0.0001 NTU up to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU; 0.01 NTU at 100.00
    NTU
    Response Time:Initial response in 1 minute, 15 seconds for a full-scale step change
    Sample Flow Rate:200 to 750 mL/minute (3.1 to 11.9 gal/hour)
    Sample Input:1/4 inch NPT female, 1/4 inch compression fitting (provided)
    Sample Output:1/2 inch NPT female, 1/2 inch hose barb (provided)
    Sample Temperature:0 to 70 °C (32 to 158 °F)
    Ship Wt. (lbs):1720E Turbidimeter and sc200 Controller: 
    8.2 kg (18.1 lbs.)
    Signal average time:User selectable from 0, 6, 30, 60, 90 seconds
    Storage Conditions:-20 to 60 °C (-4 to 140 °F)