Altitude:< 2000 m
Analog Outputs:Two (Five with optional expansion module) 0/4 to 20 mA isolated current outputs, max. 550 Ω, Accuracy: ±0.1% of FS (20 mA) at 25 °C, ±0.5% of FS over -20 °C to 60 °C range
Analogue Output Functional Mode:Linear, Logarithmic, Bi-linear, PID
Calibration Method:a) Zero: electrically, with hydrazine-free water or with optional zero cartridge
b) Slope: using a laboratory reference value (e.g. LCW025)
Communication:MODBUS RS232/RS485, PROFIBUS DPV1, HART optional
Communication Capabilities:Modbus RS232/RS485
Conduit Openings:1/2 " NPT Conduit
Connection Drain Line:1/4" OD Tubing
Connections:3/8" OD Tubing (Tubing must not exceed 4 feet and must drain straight down)
Dimensions Metric (H x W X D):817 mm x 300 mm x 228 mm
Display:Graphic dot matrix LCD with LED backlighting, transreflective
Display Resolution:240 x 160 pixels
Display Size:48 x 68 mm
Electrical Certifications:EMC
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Safety
CAN/CSA C22.2 No. 61010-1
cETLus safety mark for:
- General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Enclosure Rating:NEMA 4X/IP66
Flow:166 to 250 mL/min (10 to 15 L/h) recommended
Lower Limit of Detection (LOD):Drift is Negligible; 1 ppb
Maintenance Interval:Monthly: Calibration and reagent refill
Manual Languages:Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Material:Counter-Electrode: Stainless steel; Working Electrode: Platinum
Material Enclosures:Polycarbonate
Aluminium (powder coated)
Stainless Steel
Number of Channels:1
Operating Temperature Range:5 to 45 °C at 0 to 95% RH (non-condensing)
Power Requirements (Hz):50 - 60 Hz
Power Requirements (Voltage):100 - 240 VAC
Pressure Range:0.5 to 6 bar (7.2-87 psi) or 12 L/h
Range:0 to 500 ppb hydrazine; programmable
Range 2:0 to 100 ppb carbohydrazide; programmable
Reagent:Sample conditioning with diisopropylamine, ammonia, or diethylamine
Reference Type:Ag/AgCl electrode with electrolyte KCl 0.1 M
Relative Humidity:10 to 90 % Relative Humidity (Non-Condensing)
Relay: Operational Mode:Primary or secondary measurement, calculated value (dual channel only) or timer/scheduler
Relay Functions:Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
Relays:Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Repeatability:± 2 % or 1 ppb whichever is greater
Response Time T90:< 60 s
Sample Requirements:Sample needs to be free of undissolved matter.
Sample Temperature:5 to 45 °C
Security Levels:2
Storage Conditions:-20 °C to 70 °C
Temperature Compensation:5 to 45 °C
Warranty:1 year
Weight:32.15 lbs. (14.5824 kg)