YOUR SPECIALIST FOR
INSTRUMENTATION
& CONTROL
  Home > Product > Water Analysis > HACH > ULTRATURB plus sc







HACH COMPANY was established in 1947, in Ames, Iowa, USA. For more than 60 years, Hach has been the world market leader for Analytical Instrumentation for Complete Water Analysis, providing solutions for lab, process and field.


In 2002, GLI, a leading US manufacturer of process electro-chemistry analytical instruments, became part of the Hach Company family.

Water analysis has to be right. Our customers deserve complete, reliable solutions that they can be fully  confident in. Hach is your answer. We make water analysis simpler















Turbidimeter (Mid Range)

ULTRATURB plus sc


Turbidimeter (Mid Range)



Details


Applications: Drinking Water, Wastewater, Industrial Water.

Turbidities in the range of FNU (NTU) 0.0001-1,000 are measured and displayed alphanumerically using a controller. This high resolution facilitates precision differentiation, even for ultra clear liquids. A pulsed, long-life IR radiation source (LED) ensures that costs are minimized. Designed in accordance with DIN EN ISO 7027 and permanently calibrated before leaving the factory. All key data is set to practical standard values. The instruments are immediately ready for use after connection to the power supply and the supply of sample water. The measuring range and all data outputs are modified to suit requirements using menus on the controller. All optical and electronic assemblies are installed in housings that are physically very strong and proof against water jets. The measuring chamber also has automatic wiper cleaning.

  • Wide measuring range-0 to 1,000 NTU
  • Self-cleaning Sample Chamber Option
  • Stable/long lasting light source with IR ratio technology
  • Ratio Methodology Detects Sample Chamber Fouling
  • Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controller

Specifications

Cable Length:5 m
Dimensions Metric (H x W X D):250 mm x 240 mm x 110 mm
Display:Display provided through connection with Hach sc Controller
Enclosure Rating:IP65
Inlet:13 mm ID Tubing or fixed connection (PVC system component pipework)
Material:Housing: ASA; Window: Quartz; Wiper Blade: Silicon; Wiper Shaft: stainless steel 1.4571
Measurement Method:90° infrared pulse scattered light technique in accordance with DIN EN ISO 7027
Operating Temperature Range:2 to 40 °C (36 to 104 °F)
Outlet:13 mm ID Tubing or fixed connection (PVC system component pipework)
Range:0.0001 to 1000 NTU (TE/F, NTU, FNU) can be programmed as required (0.001-250 EBC = 2500 ppm SiO2) (Note: 1 FNU is equivalent to 1 NTU)
Reproducibility:±0.003 FNU or ±0.5% of reading (0 to 2 FNU)
Resolution:0.0001 to 0.9999 FNU
1.000 to 9.999 FNU
100 to 1000 FNU
10 to 99.99 FNU
Response Time:1 to 60 s (programmable)
Sample Flow Rate:0.2 L/min to 1 L/min
Sample Pressure:Max 6 bar at 20ºC (87 psi, 68ºF)
Sample Temperature:Max. 50 °C (122 °F)
Variation Coefficient:1 % in accordance with DIN 38402
Verification:StablCal or dry standard CVM module
Wavelength:860 nm infrared light
Weight:Approx. 3.31 lbs. (1.5 kg)